1-8 Instrument description
Status switch panel
Status switch panel
Use the three-position status switch in conjunction with the MassLynx
software to control the instrument. Two light-emitting diodes (LEDs) are
associated with the switch.
Analyser Fine The fine pumps for the analyzer region.
Rotary Pumps
Analyser The fore pump for the analyzer region.
Inlets 1 The inlets fore pump, used to pump the gas inlets,
valves, and interface regions.
Source The fore pump for the source region.
Source Fine The fine pump for the source region.
Baking The flight tube baking transformer.
Heaters The GC interface heaters and gas valve controllers.
Miniature circuit breakers (Continued)
MCB Supplies
VACUUM
OPERATE
STANDBY
ISOLATE
OPERATE