EasyManua.ls Logo

Waters Autospec Premier - Manually Correcting a Calibration

Waters Autospec Premier
354 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Manually correcting a calibration 5-7
Calibration report window
Manually correcting a calibration
See “Editing the calibration parameters” on page 5-9.
Sometimes you must manually adjust some of the matched peaks during the
calibration process. These mismatched peaks are usually apparent by the
larger-than-normal residual errors they leave. For example, in the following
figure, there are two mismatched peaks with larger residuals (highlighted).
Caution: If you manually correct a calibration but open and close the
Calibration Parameters dialog box with the “Perform auto peak
matching option” selected, the spectrum is automatically recalibrated;
any manual corrections are lost.

Table of Contents

Related product manuals