3-6 Scanning techniques
To copy an existing function in the list
1. Click on the function.
2. Click Edit > Copy.
3. Click Edit > Paste.
4. Modify the parameters as required.
To delete a function from the list:
1. Click on the function.
2. Click .
Alternative: Click Edit > Delete, or press the Del key.
To change the order of functions in the list
Rule: Functions appear in ascending start and end time order; this order;
which you cannot change. However, you can change the order of those
functions with identical start and end times.
1. Click on the required function.
2. Repeatedly click or until the function is in the desired position.
Setting the maximum retention time
To set the maximum retention (total tun) time
1. In the Total Run Time text box, enter the required value.
2. Click .
Note: The ratio of the defined functions is maintained. For example, if
two functions are defined, one from 0 to 5 minutes, and the other 5 to
10 minutes, then a total run time of 10 minutes appears automatically.
If the total run time value is changed to 20, the first function now runs
from 0 to 10 minutes, and the second from 10 to 20 minutes.