6-12 Inlets and probes
Tip: The source side lock isolation valve is open when its handle is in line
with the probe handle.
9. Close the source lock pump valve.
Tip: The probe assembly now appears, as in the following figure.
10. Rotate the probe handle so that its guide slot is correctly oriented with
the guide pin.