EasyManua.ls Logo

Waters Autospec Premier - Page 284

Waters Autospec Premier
354 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
D-10 The MassLynx Tune window
To perform automatic baking at the same time each day
1. Select Timer.
2. Select Daily.
3. Specify the Start and End times (using 24-hour notation) between which
automatic baking takes place.
Recommendation: Specify that automatic baking takes place overnight
on a daily basis, if no overnight (Night Time Process) acquisitions are
being performed.
4. Click Close.
To perform automatic baking at specified times on a weekly basis
1. Select Timer.
2. Select Weekly.
3. Select the days on which baking is to be performed.
4. Specify the Start and End times (using 24 hour notation) between which
automatic baking takes place.
5. Click Close.
To perform manual baking
1. Clear Timer.
2. Specify the required time for which baking is to take place in the Bake
for....hours box.
3. Select Baking on.
Tip: You can switch baking off at any time by selecting Baking off.
Caution: Automatic baking overrides any data acquisition that is in
progress; data are lost.
Caution: Automatic baking overrides any data acquisition that is in
progress; data are lost.