Low resolution tuning 2-5
Tune window with initial nitrogen peak
To calibrate the mass scale
1. In the Tune window, click Options > Instrument Setup.
2. On the Limits tab, drag the Magnet Limit slider until the peak is
approximately centered at mass 28.