2-12 Tuning the instrument
Flat-topped peak
6. Set the collector slit to optimize the ion beam width and shape.
Tip: Optimum transmission is achieved when the width of the collector
slit is equal to the ion beam width. When this is the case, the ion beam is
said to be triangulated.