Agilent 4155C/4156C SCPI Command Reference, Edition 1 5-75
SCPI Commands
:MMEMory:STORe:SSHeet:LINDex
Query response SPAC | TAB | COMM <newline><^END>
Example OUTPUT @Hp4155;":MMEM:STOR:SSH:DEL TAB"
:MMEMory:STORe:SSHeet:LINDex
This command specifies the range of measurement results to store in the spreadsheet
file that is created by :MMEMory:STORe:SSHeet command.
This command changes the value of :HCOPy:LINDex.
At *RST, this value is (1,MAX).
Syntax :MMEMory:STORe:SSHeet:LINDex first | MINimum | MAXimum,
last | MINimum | MAXimum
Parameter
Query response first,last <newline><^END>
The data type of first and last are NR1 response data.
Example OUTPUT @Hp4155;":MMEM:STOR:SSH:LIND 1,MAX"
OUTPUT @Hp4155;":MMEM:STOR:SSH:LIND?"
ENTER @Hp4155;A,B
See also :HCOPy:LINDex
Parameter Type Explanation
first numeric first index number to output
last numeric last index number to output