Agilent 4155C/4156C SCPI Command Reference, Edition 1 5-193
SCPI Commands
:PAGE:MEASure:QSCV:DELay
:PAGE:MEASure:QSCV:DELay
This command sets the DELAY TIME for Quasi-static CV measurements. The
delay time is the time from the start of each sweep step to the start of the
measurement. This command also has a query form.
At *RST, this value is set to 0.
Syntax :PAGE:MEASure:QSCV:DELay delay_time
:PAGE:MEASure:QSCV:DELay?
Parameter
Query response delay_time <newline><^END>
delay_time is an NR3 response data type.
Example OUTPUT @Hp4155;":PAGE:MEAS:QSCV:DEL 1.5"
OUTPUT @Hp4155;":PAGE:MEAS:QSCV:DEL?"
ENTER @Hp4155;A
Parameter Type Explanation
delay_time numeric 0 to 65.535 (s). Resolution: 100 ms.