5-94 Agilent 4155C/4156C SCPI Command Reference, Edition 1
SCPI Commands
:PAGE:CHANnels[:CDEFinition]:VMU<n>:DCHarge
:PAGE:CHANnels[:CDEFinition]:VMU<n>:DCHarge
This command connects, or disconnects, the VMU input discharge resistor used to
prevent the VMU inputs from becoming charged when the inputs are opened. This
command also has a query form.
At *RST, this value is ON.
Syntax :PAGE:CHANnels [:CDEFinition] :VMU<n>:DCHarge OFF | ON | 0 | 1
:PAGE:CHANnels [:CDEFinition] :VMU<n>:DCHarge?
<n> must be 1 or 2.
Parameter
When an auto-calibration is executed, the setting is not changed.
Query response 0 | 1 <newline><^END>
Example OUTPUT @Hp4155;":PAGE:CHAN:VMU1:DCH ON"
The command above specifies VMU1, but it connects the discharge resistor to both
the VMU1 and VMU2 inputs.
OUTPUT @Hp4155;":PAGE:CHAN:VMU1:DCH?"
ENTER @Hp4155;A
Parameter Type Explanation
OFF or 0 Boolean Disconnects the discharge resistors from the VMU
inputs.
ON or 1 Boolean Connects the discharge resistors to the VMU inputs,
and automatically breaks the connection only when
in the measurement state.

Table of Contents

Other manuals for Agilent Technologies 4156C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4156C and is the answer not in the manual?

Agilent Technologies 4156C Specifications

General IconGeneral
BrandAgilent Technologies
Model4156C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals