:PAGE:SCONtrol:STOP
:PAGE:SCONtrol:STOP
This command stops the currently executing operation (SWEEP, SAMPLING,
STRESS, or KNOB SWEEP).
Executing this command is about the same as pressing Stop front-panel key.
This command does not have query form.
Syntax :PAGE:SCONtrol:STOP
Example OUTPUT @Hp4155;":PAGE:SCON:STOP"
:PAGE:SCONtrol:STRess[:STARt]
This command starts the STRess force operation.
This command changes the present display page to the STRESS: STRESS FORCE
page.
Executing this command is about the same as pressing Stress front-panel key.
This command does not have query form.
Syntax :PAGE:SCONtrol:STRess [ :STARt]
Example OUTPUT @Hp4155;":PAGE:SCON:STR"
:PAGE:SCONtrol:TRIGger:INPut?
This command waits for the external trigger input and returns 1 just after detecting
trigger. This command has query form only.
Syntax
Parameter
:PAGE:SCONtrol:TRIGger:INPut? POSitive | NEGative | EITHer
Query response 1 <newline><^END>
Example OUTPUT @Hp4155;":PAGE:SCON:TRIG:INP? POS"
ENTER @Hp4155;A
Parameter Type Explanation
POSitive character positive slope is regarded as the trigger
NEGative character negative slope is regarded as the trigger
EITHer character positive or negative slope is regarded as the trigger

Table of Contents

Other manuals for Agilent Technologies 4156C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4156C and is the answer not in the manual?

Agilent Technologies 4156C Specifications

General IconGeneral
BrandAgilent Technologies
Model4156C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals