Agilent 4155C/4156C SCPI Command Reference, Edition 1 5-267
SCPI Commands
:PAGE:STRess:SETup:CONStant:SMU<n>[:SOURce]
:PAGE:STRess:SETup:CONStant:SMU<n>[:SOURce]
This command sets the constant SOURCE value of SMU<n> for stress.
<n> is required to specify SMU number. Valid SMU numbers are SMU1 through
SMU6, depending on the configuration.
The mode of the specified SMU must be V or I.
At *RST, this value is:
Syntax :PAGE:STRess:SETup:CONStant:SMU<n> [ :SOURce] source | MINimum |
MAXimum
Parameter
Query response source <newline><^END>
source is NR3 response data type.
Example OUTPUT @Hp4155;":PAGE:STR:SET:CONS:SMU1 10"
OUTPUT @Hp4155;":PAGE:STR:SET:CONS:SMU1 MAX"
OUTPUT @Hp4155;":PAGE:STR:SET:CONS:SMU1?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:STR:SET:CONS:SMU1? MAX"
ENTER @Hp4155;A
SMU SOURCE
SMU1 not defined
SMU2 to SMU4 0 V
Parameter Type Explanation
source numeric -200 to 200 V or -1 to 1 A. The range of this value
depends on the type of SMU.

Table of Contents

Other manuals for Agilent Technologies 4156C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4156C and is the answer not in the manual?

Agilent Technologies 4156C Specifications

General IconGeneral
BrandAgilent Technologies
Model4156C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals