5-274 Agilent 4155C/4156C SCPI Command Reference, Edition 1
SCPI Commands
:PAGE:STRess:SETup:PULSe:PGU<n>:LEADing
:PAGE:STRess:SETup:PULSe:PGU<n>:LEADing
This command sets the leading-edge transition time (LEADING TIME) of PGU<n>
for stress.
<n> is required to specify PGU number. Valid PGU numbers are PGU1 through
PGU2.
The mode of the specified PGU must be VPULse.
At *RST, this value is 100 ns.
Syntax :PAGE:STRess:SETup:PULSe:PGU<n>:LEADing leading_time | MINimum |
MAXimum
Parameter
Query response leading_time <newline><^END>
leading_time is NR3 response data type.
Example OUTPUT @Hp4155;":PAGE:STR:SET:PULS:PGU1:LEAD 0.001"
OUTPUT @Hp4155;":PAGE:STR:SET:PULS:PGU1:LEAD MAX"
OUTPUT @Hp4155;":PAGE:STR:SET:PULS:PGU1:LEAD?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:STR:SET:PULS:PGU1:LEAD? MAX"
ENTER @Hp4155;A
Parameter Type Explanation
leading_time numeric 100E-9 to 10E-3 (s)