5-266 Agilent 4155C/4156C SCPI Command Reference, Edition 1
SCPI Commands
:PAGE:STRess:SETup:CONStant:SMU<n>:COMPliance
:PAGE:STRess:SETup:CONStant:SMU<n>:COMPlian
ce
This command sets the constant COMPLIANCE value of SMU<n> for stress.
<n> is required to specify SMU number. Valid SMU numbers are SMU1 through
SMU6, depending on the configuration.
The mode of the specified SMU must be V or I.
At *RST, this value is:
Syntax :PAGE:STRess:SETup:CONStant:SMU<n>:COMPliance compliance | MINmum |
MAXimum
Parameter
Query response compliance <newline><^END>
compliance is NR3 response data type.
Example OUTPUT @Hp4155;":PAGE:STR:SET:CONS:SMU1:COMP 0.1"
OUTPUT @Hp4155;":PAGE:STR:SET:CONS:SMU1:COMP MAX"
OUTPUT @Hp4155;":PAGE:STR:SET:CONS:SMU1:COMP?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:STR:SET:CONS:SMU1:COMP? MAX"
ENTER @Hp4155;A
SMU COMPLIANCE
SMU1 not defined
SMU2 to SMU4 100 mA
SMU5 and SMU6 not defined
Parameter Type Explanation
compliance numeric -200 to 200 V or -1 to 1 A. The range of this value
depends on the type of SMU.