5-296 Agilent 4155C/4156C SCPI Command Reference, Edition 1
SCPI Commands
:STATus:QUEStionable:CONDition?
The preset value of each register as follows:
:STATus:QUEStionable:CONDition?
This command returns the present status of the Questionable Status "CONDITION"
register.
Reading this register does not clear it.
This command has query form only.
Syntax :STATus:QUEStionable:CONDition?
Query response register <newline><^END>
register is a decimal integer value, which is the sum of the binary-weighted values,
in NR1 response data format.
Example OUTPUT @Hp4155;":STAT:QUES:COND?"
ENTER @Hp4155;A
Register Filter/Enable PREset value
OPERation ENABle 0s
PTR 1s
NTR 0s
QUEStionable ENABle 0s
PTR 1s
NTR 0s
EMERgency ENABle 1s
PTR 1s
NTR 0s