5-172 Agilent 4155C/4156C SCPI Command Reference, Edition 1
SCPI Commands
:PAGE:MEASure:MSETup:WTIMe
:PAGE:MEASure:MSETup:WTIMe
This command sets the WAIT TIME multiplier for the measurement. The default
wait time is multiplied by this value.
Actual Wait Time = Default Wait Time ´ wait time
At *RST, this value is 1.
Syntax :PAGE:MEASure:MSETup:WTIMe wait_time | MINimum | MAXimum
Parameter
Query response wait_time <newline><^END>
wait_time is NR3 response data type.
Example OUTPUT @Hp4155;":PAGE:MEAS:MSET:WTIM 1.0"
OUTPUT @Hp4155;":PAGE:MEAS:MSET:WTIM MAX"
OUTPUT @Hp4155;":PAGE:MEAS:MSET:WTIM?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:MEAS:MSET:WTIM? MAX"
ENTER @Hp4155;A
:PAGE:MEASure:MSETup:ZCANcel:SMU<n>?|:VMU
<n>?
This command query returns whether the ZERO OFFSET CANCEL mode is
enabled for the specified unit.
If :PAGE:MEAS:MSET:ZCANcel[:STATe] is OFF, this query always returns
0.
<n> is required to specify SMU or VMU number. Valid numbers are SMU1 through
SMU6 (depending on the configuration) and VMU1 through VMU2.
The specified SMU or VMU must not be DISable.
This command has query form only.
Syntax :PAGE:MEASure:MSETup:ZCANcel:SMU<n>? | :VMU<n>?
Parameter Type Explanation
wait_time numeric 0 to 10 with resolution 0.1.