Agilent 4155C/4156C SCPI Command Reference, Edition 1 5-211
SCPI Commands
:PAGE:MEASure:SAMPling:FILTer
:PAGE:MEASure:SAMPling:FILTer
This command sets the SMU output filter state for sampling measurement.
Syntax :PAGE:MEASure:SAMPling:FILTer OFF | ON | 0 | 1
Parameter
Query response 0 | 1 <newline><^END>
Example OUTPUT @Hp4155;":PAGE:MEASure:SAMPling:FILT ON"
OUTPUT @Hp4155;":PAGE:MEASure:SAMPling:FILT?"
ENTER @Hp4155;A
:PAGE:MEASure:SAMPling:HTIMe
This command sets the HOLD TIME of sampling measurement.
Syntax :PAGE:MEASure:SAMPling:HTIMe hold_time | MINimum | MAXimum
Parameter
Query response hold_time <newline><^END>
hold_time is NR3 response data type.
Example OUTPUT @Hp4155;":PAGE:MEAS:SAMP:HTIM 1.5"
OUTPUT @Hp4155;":PAGE:MEAS:SAMP:HTIM MAX"
OUTPUT @Hp4155;":PAGE:MEAS:SAMP:HTIM?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:MEAS:SAMP:HTIM? MAX"
ENTER @Hp4155;A
Parameter Type Explanation
OFF or 0 boolean Filter off
ON or 1 boolean Filter on
Parameter Type Explanation
hold_time numeric 3E-2 to 655.35 (s). Resolution: 100 ms