5-48 Agilent 4155C/4156C SCPI Command Reference, Edition 1
SCPI Commands
:HCOPy:ITEM[:WINDow]:TEXT2:STATe
:HCOPy:ITEM[:WINDow]:TEXT2:STATe
This command sets whether the GRAPH AXIS TEXT (names, units, and scale of
the graph axis) should be plotted or printed when the :HCOPy:IMMediate command
or :HCOPy:DATA? query is sent.
At *RST, the value of this parameter is ON.
This command does not comply with SCPI.
Syntax :HCOPy:ITEM [ :WINDow] :TEXT2:STATe OFF | ON | 0 | 1
Parameter
This command has meaning only when :HCOPy:DEVice:LANGuage is set to
HP-GL. If set to PCL, GRAPH AXIS TEXT is always printed or plotted.
Query response 0 | 1 <newline><^END>
Example OUTPUT @Hp4155;":HCOP:ITEM:TEXT2:STAT ON"
OUTPUT @Hp4155;":HCOP:ITEM:TEXT2:STAT?"
ENTER @Hp4155;A
:HCOPy:ITEM[:WINDow]:TEXT3:STATe
This command sets whether the measurement setup data should be plotted or printed
when the :HCOPy:IMMediate command or :HCOPy:DATA? query is sent to output
graphics results.
At *RST, the value of this parameter is OFF.
This command does not comply with SCPI.
Syntax :HCOPy:ITEM [ :WINDow] :TEXT3:STATe OFF | ON | 0 | 1
Parameter Type Explanation
OFF or 0 boolean GRAPH AXIS TEXT is not plotted or printed.
ON or 1 boolean GRAPH AXIS TEXT is plotted or printed.