EasyManuals Logo

Agilent Technologies 4156C Command Reference

Agilent Technologies 4156C
424 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #290 background imageLoading...
Page #290 background image
5-208 Agilent 4155C/4156C SCPI Command Reference, Edition 1
SCPI Commands
:PAGE:MEASure:SAMPling:CONStant:SMU<n>:COMPliance
:PAGE:MEASure:SAMPling:CONStant:SMU<n>:CO
MPliance
This command sets the constant COMPLIANCE value of SMU<n> for the sampling
measurement.
<n> is required to specify SMU number. Valid SMU numbers are SMU1 through
SMU6, depending on the configuration.
The mode of the specified SMU must be V or I.
Syntax :PAGE:MEASure:SAMPling:CONStant:SMU<n>:COMPliance compliance |
MINimum | MAXimum
Parameter
Query response compliance <newline><^END>
compliance is NR3 response data type.
Example OUTPUT @Hp4155;":PAGE:MEAS:SAMP:CONS:SMU1:COMP 0.1"
OUTPUT @Hp4155;":PAGE:MEAS:SAMP:CONS:SMU1:COMP MAX"
OUTPUT @Hp4155;":PAGE:MEAS:SAMP:CONS:SMU1:COMP?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:MEAS:SAMP:CONS:SMU1:COMP? MAX"
ENTER @Hp4155;A
Parameter Type Explanation
compliance numeric -200 to 200 V or -1 to 1 A. The range of this value
depends on the type of SMU.

Table of Contents

Other manuals for Agilent Technologies 4156C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4156C and is the answer not in the manual?

Agilent Technologies 4156C Specifications

General IconGeneral
BrandAgilent Technologies
Model4156C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals