Agilent 4155C/4156C SCPI Command Reference, Edition 1 3-15
Status Reporting Structure
Measurement/Stress Status Register
Measurement/Stress Status Register
This register consists of EVENT and event enable (MASK) registers.
You enable the desired bits of the EVENT register by using the MASK register. 1
enables, 0 masks the corresponding bit of EVENT register. The MASK register is
set by the :STATus:MEASurement:ENABle command. See STATus Subsystem in
Chapter 5 for details about this command.
Figure 3-9 Measurement/Stress Status Register
This register is not a standard SCPI register.
Table 3-6 Measurement/Stress Status Register
Bit Definition Explanation
0 A/D Overflow This bit is set if A/D converter overflows.
1 Oscillation Status This bit is set if SMU oscillates.
2 NOT USED This bit is always zero.
3 Compliance Status This bit is set if SMU reaches compliance.
4 to 7 NOT USED These bits are always zero.
8 PGU Status This bit is set if PGU reaches the current limit.
9 to 15 NOT USED These bits are always zero.

Table of Contents

Other manuals for Agilent Technologies 4156C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4156C and is the answer not in the manual?

Agilent Technologies 4156C Specifications

General IconGeneral
BrandAgilent Technologies
Model4156C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals