Agilent 4155C/4156C SCPI Command Reference, Edition 1 5-255
SCPI Commands
:PAGE:STRess[:CDEFinition]:SELector<n>:MEASure
Query response name <newline><^END>
name is string response data, but does not contain double quote characters at the
beginning and end of the string.
Example OUTPUT @Hp4155;":PAGE:STR:PGU1:NAME'VD'"
OUTPUT @Hp4155;":PAGE:STR:PGU1:NAME?"
ENTER @Hp4155;A$
:PAGE:STRess[:CDEFinition]:SELector<n>:MEASure
This command selects the relay connection mode of Agilent 16440A SMU/PGU
selector for measurement state.
<n> is required to specify the SMU/PGU selector number. Valid SMU/PGU selector
numbers are SELector1 through SELector4.
At *RST, this value is OPEN.
Syntax :PAGE:STRess [ :CDEFinition] :SELector<n>:MEASure SMU | PGU | OPEN |
POPen
Parameter
Query response SMU | PGU | OPEN | POP <newline><^END>
Example OUTPUT @Hp4155;":PAGE:STR:SEL1:MEAS SMU"
OUTPUT @Hp4155;":PAGE:STR:SEL1:MEAS?"
ENTER @Hp4155;A$
Parameter Type Explanation
SMU character connect to SMU port
PGU character connect to PGU port
OPEN character disconnect all
POPen character disconnect all, but disconnect PGU port with
semiconductor switch, not normal switch. This
parameter is valid only for selector 1 and 3.

Table of Contents

Other manuals for Agilent Technologies 4156C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4156C and is the answer not in the manual?

Agilent Technologies 4156C Specifications

General IconGeneral
BrandAgilent Technologies
Model4156C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals