6-8 Agilent 4155C/4156C SCPI Command Reference, Edition 1
Error Messages
Execution Error
Execution Error
The 4155C/4156C reports -2XX errors when it is unable to perform a valid
programming command.
Error -200 Execution error
Generic execution error for the 4155C/4156C that cannot be determined more
specifically.
Error -201 Invalid while in local
A command is not executable while the 4155C/4156C is in local mode due to a Hard
Local Control.
Error -202 Settings lost due to rtl
A setting associated with a Hard Local Control was lost when the 4155C/4156C
changed to Local State (LOCS) from Remote State (REMS) or to Local with
Lockout State (LWLS) from Remote with Lockout State (RWLS).
Error -210 Trigger error
Error -211 Trigger ignored
A GET(Group Execution Trigger), *TRG, or triggering signal was received and
recognized by the 4155C/4156C but was ignored because of timing considerations;
for example, the 4155C/4156C was not ready to respond.
Error -214 Trigger deadlock
The trigger source for the initiation of a measurement is set to GET (Group
Execution Trigger) and subsequent measurement query is received. The
measurement cannot be started until a GET is received.
Error -220 Parameter error
A parameter related error occurred and the 4155C/4156C cannot determine the more
specific errors -221 through -224.
Error -221 Settings conflict
A specified parameter setting could not be executed due to the present device state.