Agilent 4155C/4156C SCPI Command Reference, Edition 1 6-9
Error Messages
Execution Error
Error -222 Data out of range
Interpreted value of the program was out of range as defined by the 4155C/4156C.
Error -223 Too much data
Too many parameters for the 4155C/4156C was received.
Error -224 Illegal parameter value
Illegal parameter value was received.
Error -230 Data corrupt or stale
Possibly invalid data; new reading started but not completed since last access.
Error -231 Data questionable
Measurement accuracy is suspect.
Error -240 Hardware error
A hardware problem in the 4155C/4156C. This error message is reported if the
4155C/4156C cannot detect the more specific error -241.
Error -241 Hardware missing
A program command or query could not be executed because of missing hardware;
for example, an option was not installed.
Error -250 Mass storage error
A mass storage error occurred. This error message is reported if the 4155C/4156C
cannot determine the more specific errors -251 through -258.
Error -251 Missing mass storage
A program command or query could not be executed because of missing mass
storage.
Error -252 Missing media
A program command or query could not be executed because of a missing media.

Table of Contents

Other manuals for Agilent Technologies 4156C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4156C and is the answer not in the manual?

Agilent Technologies 4156C Specifications

General IconGeneral
BrandAgilent Technologies
Model4156C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals