5-210 Agilent 4155C/4156C SCPI Command Reference, Edition 1
SCPI Commands
:PAGE:MEASure:SAMPling:CONStant:VSU<n>[:SOURce]
:PAGE:MEASure:SAMPling:CONStant:VSU<n>[:SOU
Rce]
This command sets the constant SOURCE value of VSU<n> for the sampling
measurement.
<n> is required to specify VSU number. Valid VSU numbers are VSU1 through
VSU2.
The specified VSU must not be DISable.
Syntax :PAGE:MEASure:SAMPling:CONStant:VSU<n> [ :SOURce] source | MINimum |
MAXimum
Parameter
Query response source <newline><^END>
source is NR3 response data type.
Example OUTPUT @Hp4155;":PAGE:MEAS:SAMP:CONS:VSU1 10"
OUTPUT @Hp4155;":PAGE:MEAS:SAMP:CONS:VSU1 MAX"
OUTPUT @Hp4155;":PAGE:MEAS:SAMP:CONS:VSU1?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:MEAS:SAMP:CONS:VSU1? MAX"
ENTER @Hp4155;A
Parameter Type Explanation
source numeric -20 V to 20 V