Agilent 4155C/4156C SCPI Command Reference, Edition 1 5-273
SCPI Commands
:PAGE:STRess:SETup:PULSe:PGU<n>:IMPedance
Query response delay_time <newline><^END>
delay_time is NR3 response data type.
Example OUTPUT @Hp4155;":PAGE:STR:SET:PULS:PGU1:DEL 1"
OUTPUT @Hp4155;":PAGE:STR:SET:PULS:PGU1:DEL MAX"
OUTPUT @Hp4155;":PAGE:STR:SET:PULS:PGU1:DEL?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:STR:SET:PULS:PGU1:DEL? MAX"
ENTER @Hp4155;A
:PAGE:STRess:SETup:PULSe:PGU<n>:IMPedance
This command sets the output IMPEDANCE of PGU<n> for stress.
<n> is required to specify PGU number. Valid PGU numbers are PGU1 through
PGU2.
The mode of the specified PGU must be VPULse.
At *RST, this value is LOW.
Syntax :PAGE:STRess:SETup:PULSe:PGU<n>:IMPedance LOW | R50
Parameter
Query response LOW | R50 <newline><^END>
Example OUTPUT @Hp4155;":PAGE:STR:SET:PULS:PGU1:IMP LOW"
OUTPUT @Hp4155;":PAGE:STR:SET:PULS:PGU1:IMP?"
ENTER @Hp4155;A$
Parameter Type Explanation
LOW character low impedance
R50 character 50 W

Table of Contents

Other manuals for Agilent Technologies 4156C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4156C and is the answer not in the manual?

Agilent Technologies 4156C Specifications

General IconGeneral
BrandAgilent Technologies
Model4156C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals