Agilent 4155C/4156C SCPI Command Reference, Edition 1 5-185
SCPI Commands
:PAGE:MEASure:PGUSetup:PULSe:PGU<n>:PERiod
:PAGE:MEASure:PGUSetup:PULSe:PGU<n>:PERiod
This command sets the pulse PERIOD of PGU<n>.
<n> is required to specify PGU number. Valid PGU numbers are PGU1 through
PGU2.
PGU1:PERiod and PGU2:PERiod are always set to the same value.
You use this command only if the specified PGU is set to VPULse mode (not V) by
the :PAGE:CHAN:PGU<n>:MODE command.
At *RST, this value is 10 ms.
Syntax :PAGE:MEASure:PGUSetup:PULSe:PGU<n>:PERiod period | MINimum |
MAXimum
Parameter
Query response period <newline><^END>
period is NR3 response data type.
Example OUTPUT @Hp4155;":PAGE:MEAS:PGUS:PULS:PGU1:PER 0.01"
OUTPUT @Hp4155;":PAGE:MEAS:PGUS:PULS:PGU1:PER MIN"
OUTPUT @Hp4155;":PAGE:MEAS:PGUS:PULS:PGU1:PER?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:MEAS:PGUS:PULS:PGU1:PER? MIN"
ENTER @Hp4155;A
Parameter Type Explanation
period numeric 2E-6 to 10 (s)