5-272 Agilent 4155C/4156C SCPI Command Reference, Edition 1
SCPI Commands
:PAGE:STRess:SETup:PULSe:PGU<n>:BASE
:PAGE:STRess:SETup:PULSe:PGU<n>:BASE
This command sets the BASE VALUE of PGU<n> for stress.
<n> is required to specify PGU number. Valid PGU numbers are PGU1 through
PGU2.
The mode of the specified PGU must be VPULse.
At *RST, this value is 0.
Syntax :PAGE:STRess:SETup:PULSe:PGU<n>:BASE base | MINimum | MAXimum
Parameter
Query response base <newline><^END>
base is NR3 response data type.
Example OUTPUT @Hp4155;":PAGE:STR:SET:PULS:PGU1:BASE 1"
OUTPUT @Hp4155;":PAGE:STR:SET:PULS:PGU1:BASE?"
ENTER @Hp4155;A
:PAGE:STRess:SETup:PULSe:PGU<n>:DELay
This command sets the DELAY TIME of PGU<n> for stress.
<n> is required to specify PGU number. Valid PGU numbers are PGU1 through
PGU2.
The mode of the specified PGU must be VPULse.
At *RST, this value is 0.
Syntax :PAGE:STRess:SETup:PULSe:PGU<n>:DELay delay_time | MINimum |
MAXimum
Parameter
Parameter Type Explanation
base numeric -40 to 40 V
Parameter Type Explanation
delay_time numeric 0 to 10 (s)