EasyManuals Logo

Agilent Technologies 4156C Command Reference

Agilent Technologies 4156C
424 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #354 background imageLoading...
Page #354 background image
5-272 Agilent 4155C/4156C SCPI Command Reference, Edition 1
SCPI Commands
:PAGE:STRess:SETup:PULSe:PGU<n>:BASE
:PAGE:STRess:SETup:PULSe:PGU<n>:BASE
This command sets the BASE VALUE of PGU<n> for stress.
<n> is required to specify PGU number. Valid PGU numbers are PGU1 through
PGU2.
The mode of the specified PGU must be VPULse.
At *RST, this value is 0.
Syntax :PAGE:STRess:SETup:PULSe:PGU<n>:BASE base | MINimum | MAXimum
Parameter
Query response base <newline><^END>
base is NR3 response data type.
Example OUTPUT @Hp4155;":PAGE:STR:SET:PULS:PGU1:BASE 1"
OUTPUT @Hp4155;":PAGE:STR:SET:PULS:PGU1:BASE?"
ENTER @Hp4155;A
:PAGE:STRess:SETup:PULSe:PGU<n>:DELay
This command sets the DELAY TIME of PGU<n> for stress.
<n> is required to specify PGU number. Valid PGU numbers are PGU1 through
PGU2.
The mode of the specified PGU must be VPULse.
At *RST, this value is 0.
Syntax :PAGE:STRess:SETup:PULSe:PGU<n>:DELay delay_time | MINimum |
MAXimum
Parameter
Parameter Type Explanation
base numeric -40 to 40 V
Parameter Type Explanation
delay_time numeric 0 to 10 (s)

Table of Contents

Other manuals for Agilent Technologies 4156C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4156C and is the answer not in the manual?

Agilent Technologies 4156C Specifications

General IconGeneral
BrandAgilent Technologies
Model4156C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals