5-258 Agilent 4155C/4156C SCPI Command Reference, Edition 1
SCPI Commands
:PAGE:STRess[:CDEFinition]:SMU<n>:MODE
:PAGE:STRess[:CDEFinition]:SMU<n>:MODE
This command sets the output MODE of SMU<n> for stress.
<n> is required to specify SMU number. Valid SMU numbers are SMU1 through
SMU6, depending on the configuration.
At *RST, this value is:
Syntax :PAGE:STRess [ :CDEFinition] :SMU<n>:MODE V | I | COMMon
Parameter
Query response V | I | COMM | DIS <newline><^END>
If SMU is not used, the response data may be DIS (DISable).
Example OUTPUT @Hp4155;":PAGE:STR:SMU1:MODE V"
OUTPUT @Hp4155;":PAGE:STR:SMU1:MODE?"
ENTER @Hp4155;A$
SMU OUTPUT MODE
SMU1 COMMon
SMU2 to SMU4 V
SMU5 and SMU6 not defined.
Parameter Type Explanation
V character voltage output mode
I character current output mode
COMMon character common

Table of Contents