EasyManuals Logo

Agilent Technologies 4156C Command Reference

Agilent Technologies 4156C
424 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #262 background imageLoading...
Page #262 background image
5-180 Agilent 4155C/4156C SCPI Command Reference, Edition 1
SCPI Commands
:PAGE:MEASure:PGUSetup:PULSe:PGU<n>:COUNt
:PAGE:MEASure:PGUSetup:PULSe:PGU<n>:COUNt
This command sets the output PULSE COUNT of PGU<n>.
<n> is required to specify PGU number. Valid PGU numbers are PGU1 through
PGU2.
PGU1:COUNt and PGU2:COUNt are always set to the same value.
You use this command only if the specified PGU is set to VPULse mode (not V) by
the :PAGE:CHAN:PGU<n>:MODE command.
At *RST, this value is 0.
Syntax :PAGE:MEASure:PGUSetup:PULSe:PGU<n>:COUNt count | MINimum |
MAXimum
Parameter
Query response count <newline><^END>
count is NR1 response data type.
Example OUTPUT @Hp4155;":PAGE:MEAS:PGUS:PULS:PGU1:COUNT 10"
OUTPUT @Hp4155;":PAGE:MEAS:PGUS:PULS:PGU1:COUNT MAX"
OUTPUT @Hp4155;":PAGE:MEAS:PGUS:PULS:PGU1:COUNT?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:MEAS:PGUS:PULS:PGU1:COUNT? MAX"
ENTER @Hp4155;A
Parameter Type Explanation
count numeric 0 to 65535 (0 means free run)

Table of Contents

Other manuals for Agilent Technologies 4156C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4156C and is the answer not in the manual?

Agilent Technologies 4156C Specifications

General IconGeneral
BrandAgilent Technologies
Model4156C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals