5-180 Agilent 4155C/4156C SCPI Command Reference, Edition 1
SCPI Commands
:PAGE:MEASure:PGUSetup:PULSe:PGU<n>:COUNt
:PAGE:MEASure:PGUSetup:PULSe:PGU<n>:COUNt
This command sets the output PULSE COUNT of PGU<n>.
<n> is required to specify PGU number. Valid PGU numbers are PGU1 through
PGU2.
PGU1:COUNt and PGU2:COUNt are always set to the same value.
You use this command only if the specified PGU is set to VPULse mode (not V) by
the :PAGE:CHAN:PGU<n>:MODE command.
At *RST, this value is 0.
Syntax :PAGE:MEASure:PGUSetup:PULSe:PGU<n>:COUNt count | MINimum |
MAXimum
Parameter
Query response count <newline><^END>
count is NR1 response data type.
Example OUTPUT @Hp4155;":PAGE:MEAS:PGUS:PULS:PGU1:COUNT 10"
OUTPUT @Hp4155;":PAGE:MEAS:PGUS:PULS:PGU1:COUNT MAX"
OUTPUT @Hp4155;":PAGE:MEAS:PGUS:PULS:PGU1:COUNT?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:MEAS:PGUS:PULS:PGU1:COUNT? MAX"
ENTER @Hp4155;A
Parameter Type Explanation
count numeric 0 to 65535 (0 means free run)