Agilent 4155C/4156C SCPI Command Reference, Edition 1 5-121
SCPI Commands
:PAGE:DISPlay:ANALysis[:MENU]
:PAGE:DISPlay:ANALysis[:MENU]
This command changes the present display page to DISPLAY: ANALYSIS SETUP
page.
This command does not have query form.
Syntax :PAGE:DISPlay:ANALysis [ :MENU]
Example OUTPUT @Hp4155;":PAGE:DISP:ANAL"
:PAGE:DISPlay[:SETup]:DRESOL
This command changes the resolution of the measurement data displayed on the
screen. This command also has a query form.
At *RST, this value is set to set to NORM.
Syntax :PAGE:DISPlay [:SETup] :DRESOL NORMal | EXTend
:PAGE:DISPlay:SETup:DRESOL?
Parameter
Query response NORM | EXT <newline><^END>
Example OUTPUT @Hp4155;":PAGE:DISP:SET:DRESOL EXT"
OUTPUT @Hp4155;":PAGE:DISP:SET:DRESOL?"
ENTER @Hp4155;A$
Parameter Type Explanation
NORMal character The resolution depends on the measurement
resolution described in the specifications. For
example, 1 fA resolution on the 10 pA range.
EXTend character The resolution is extended up to the full scale of
the internal A/D converter. For example, 10 aA
resolution on the 10 pA range.

Table of Contents

Other manuals for Agilent Technologies 4156C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4156C and is the answer not in the manual?

Agilent Technologies 4156C Specifications

General IconGeneral
BrandAgilent Technologies
Model4156C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals