Agilent 4155C/4156C SCPI Command Reference, Edition 1 5-271
SCPI Commands
:PAGE:STRess:SETup:PCOunt
Parameter
Query response DURation | PCOunt <newline><^END>
Example OUTPUT @Hp4155;":PAGE:STR:SET:MODE DUR"
OUTPUT @Hp4155;":PAGE:STR:SET:MODE?"
ENTER @Hp4155;A$
:PAGE:STRess:SETup:PCOunt
This command sets the stress time by the number of output pulses.
If :SETup:MODE is DURation, this parameter is not used.
Syntax :PAGE:STRess:SETup:PCOunt pcount | MINimum | MAXimum
Parameter
Query response pcount <newline><^END>
pcount is NR1 response data type.
Example OUTPUT @Hp4155;":PAGE:STR:SET:PCO 10"
OUTPUT @Hp4155;":PAGE:STR:SET:PCO MAX"
OUTPUT @Hp4155;":PAGE:STR:SET:PCO?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:STR:SET:PCO? MAX"
ENTER @Hp4155;A
Parameter Type Explanation
DURation character specify the stress time by duration
PCOunt character specify the stress time by pulse count
Parameter Type Explanation
pcount numeric 0 to 65535. 0 means free run.