5-238 Agilent 4155C/4156C SCPI Command Reference, Edition 1
SCPI Commands
:PAGE:MEASure[:SWEep]:VARD:PCOMpliance
:PAGE:MEASure[:SWEep]:VARD:PCOMpliance
This command sets the power compliance value of VARD.
If the unit type of VARD is VSU, this parameter is ignored.
You use this command only if there is an SMU whose function (FCTN) is VAR1'.
Syntax :PAGE:MEASure [ :SWEep] :VARD:PCOMPliance pcompliance | MINimum |
MAXimum
Parameter
If SMU for VAR1' unit is set to VPULSE or IPULSE mode, you cannot set power
compliance for VAR1' unit.
Query response pcompliance <newline><^END>
pcompliance is NR3 response data type.
Example OUTPUT @Hp4155;":PAGE:MEAS:VARD:PCOM 0.5"
OUTPUT @Hp4155;":PAGE:MEAS:VARD:PCOM MAX"
OUTPUT @Hp4155;":PAGE:MEAS:VARD:PCOM?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:MEAS:VARD:PCOM? MAX"
ENTER @Hp4155;A
Parameter Type Explanation
pcompliance numeric 1E-3 to 20. The range of this value depends on
SMU type of VAR1'.

Table of Contents

Other manuals for Agilent Technologies 4156C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4156C and is the answer not in the manual?

Agilent Technologies 4156C Specifications

General IconGeneral
BrandAgilent Technologies
Model4156C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals