5-194 Agilent 4155C/4156C SCPI Command Reference, Edition 1
SCPI Commands
:PAGE:MEASure:QSCV:HTIMe
:PAGE:MEASure:QSCV:HTIMe
This command sets the HOLD TIME for Quasi-static CV measurements. The hold
time is the time from the start of the first sweep step to the start of the delay time.
This command also has a query form.
At *RST, this value is set to 0.
Syntax :PAGE:MEASure:QSCV:HTIMe hold_time
:PAGE:MEASure:QSCV:HTIMe?
Parameter
Query response hold_time <newline><^END>
hold_time is an NR3 response data type.
Example OUTPUT @Hp4155;":PAGE:MEAS:QSCV:HTIM 1.5"
OUTPUT @Hp4155;":PAGE:MEAS:QSCV:HTIM?"
ENTER @Hp4155;A
Parameter Type Explanation
hold_time numeric 0 to 655.35 (s). Resolution: 10 ms.