Agilent 4155C/4156C SCPI Command Reference, Edition 1 5-237
SCPI Commands
:PAGE:MEASure[:SWEep]:VARD:OFFSet
:PAGE:MEASure[:SWEep]:VARD:OFFSet
This command sets the OFFSET value of VAR1'.
For each step of sweep, the output values of VAR1' are determined by the following
equation:
VAR1’ = VAR1 ´ RATio + OFFSet
You use this command only if there is an SMU or VSU whose function (FCTN) is
VA R 1 ' .
Syntax :PAGE:MEASure [ :SWEep] :VARD:OFFSet offset | MINimum | MAXimum
Parameter
Query response offset <newline><^END>
offset is NR3 response data type.
Example OUTPUT @Hp4155;":PAGE:MEAS:VARD:OFFSET 0.5"
OUTPUT @Hp4155;":PAGE:MEAS:VARD:OFFSET?"
ENTER @Hp4155;A
See also :PAGE:MEASure[:SWEep]:VARD:RATio
Parameter Type Explanation
offset numeric -400 to 400 V or -2 to 2 A. The range of this value
depends on the unit type of VAR1'.