5-232 Agilent 4155C/4156C SCPI Command Reference, Edition 1
SCPI Commands
:PAGE:MEASure[:SWEep]:VAR2:COMPliance
:PAGE:MEASure[:SWEep]:VAR2:COMPliance
This command sets the COMPLIANCE value of VAR2.
You use this command only if there is an SMU whose function (FCTN) is VAR2.
If the unit type of VAR2 is VSU, this parameter is ignored.
At *RST, this value is 2 V.
Syntax :PAGE:MEASure [ :SWEep] :VAR2:COMPliance compliance | MINimum |
MAXimum
Parameter
Query response compliance <newline><^END>
compliance is NR3 response data type.
Example OUTPUT @Hp4155;":PAGE:MEAS:VAR2:COMP 2"
OUTPUT @Hp4155;":PAGE:MEAS:VAR2:COMP MAX"
OUTPUT @Hp4155;":PAGE:MEAS:VAR2:COMP?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:MEAS:VAR2:COMP? MAX"
ENTER @Hp4155;A
:PAGE:MEASure[:SWEep]:VAR2:PCOMpliance
This command sets the power compliance value of VAR2.
You use this command only if there is an SMU whose function (FCTN) is VAR2.
If the unit type of VAR2 is VSU, this parameter is ignored.
At *RST, this value is not defined, but the PCOMpliance:STATe is OFF.
Syntax :PAGE:MEASure [ :SWEep] :VAR2:PCOMpliance pcompliance | MINimum |
MAXimum
Parameter Type Explanation
compliance numeric -200 to 200 V or -1 to 1 A. The range of this value
depends on SMU type of VAR2.