EasyManuals Logo

Agilent Technologies 4156C Command Reference

Agilent Technologies 4156C
424 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #260 background imageLoading...
Page #260 background image
5-178 Agilent 4155C/4156C SCPI Command Reference, Edition 1
SCPI Commands
:PAGE:MEASure:PGUSetup:CONStant:PGU<n>[:SOURce]
:PAGE:MEASure:PGUSetup:CONStant:PGU<n>[:SO
URce]
This command sets the constant SOURCE value of PGU<n>.
<n> is required to specify PGU number. Valid PGU numbers are PGU1 through
PGU2.
You use this command only if the specified PGU is set to V mode (not VPULSE) by
the :PAGE:CHAN:PGU<n>:MODE command.
At *RST, this value is 0.
Syntax :PAGE:MEASure:PGUSetup:CONStant:PGU<n> [ :SOURce] source | MINImum |
MAXimum
Parameter
Query response source <newline><^END>
source is NR3 response data type.
Example OUTPUT @Hp4155;":PAGE:MEAS:PGUS:CONS:PGU1 10"
OUTPUT @Hp4155;":PAGE:MEAS:PGUS:CONS:PGU1 MAX"
OUTPUT @Hp4155;":PAGE:MEAS:PGUS:CONS:PGU1?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:MEAS:PGUS:CONS:PGU1? MAX"
ENTER @Hp4155;A
:PAGE:MEASure:PGUSetup[:MENU]
This command changes the present display page to MEASURE: PGU SETUP.
This command does not have query form.
Syntax :PAGE:MEASure:PGUSetup [ :MENU]
Example OUTPUT @Hp4155;":PAGE:MEAS:PGUS"
Parameter Type Explanation
source numeric -40 to 40 V

Table of Contents

Other manuals for Agilent Technologies 4156C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4156C and is the answer not in the manual?

Agilent Technologies 4156C Specifications

General IconGeneral
BrandAgilent Technologies
Model4156C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals