5-168 Agilent 4155C/4156C SCPI Command Reference, Edition 1
SCPI Commands
:PAGE:MEASure:MSETup:ITIMe:SHORt[:APERture]
:PAGE:MEASure:MSETup:ITIMe:SHORt[:APERture
]
This command sets the SHORT integration time in terms of the APERTURE, which
means time.
The parameter has units of seconds.
At *RST, this value is 640 ms.
Syntax :PAGE:MEASure:MSETup:ITIMe:SHORt [ :APERture] aperture | MINimum |
MAXimum
Parameter
Query response aperture <newline><^END>
aperture is NR3 response data type.
Example OUTPUT @Hp4155;":PAGE:MEAS:MSET:ITIM:SHORT 1E-3"
OUTPUT @Hp4155;":PAGE:MEAS:MSET:ITIM:SHORT MIN"
OUTPUT @Hp4155;":PAGE:MEAS:MSET:ITIM:SHORT?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:MEAS:MSET:ITIM:SHORT? MIN"
ENTER @Hp4155;A
Parameter Type Explanation
aperture numeric 8E-5 to 1.92E-3 (s)