Loading...
Agilent 4155C/4156C SCPI Command Reference, Edition 1 5-137
SCPI Commands
:PAGE:GLISt[:GRAPhics]:DISPlay:GRID
:PAGE:GLISt[:GRAPhics]:DISPlay:GRID
This command selects whether the grid of graph axis is on or off.
At *RST, this value is ON.
Syntax :PAGE:GLISt [ :GRAPhics] :DISPlay:GRID OFF | ON | 0 | 1
Parameter
Query response 0 | 1 <newline><^END>
Example OUTPUT @Hp4155;":PAGE:GLIS:DISP:GRID ON"
OUTPUT @Hp4155;":PAGE:GLIS:DISP:GRID?"
ENTER @Hp4155;A
:PAGE:GLISt[:GRAPhics]:DISPlay:LPARam
This command selects whether the line parameters (gradient and intercepts of lines)
are displayed on the graph.
At *RST, this value is ON.
Syntax :PAGE:GLISt [ :GRAPhics] :DISPlay:LPARam OFF | ON | 0 | 1
Parameter
Query response 0 | 1 <newline><^END>
Example OUTPUT @Hp4155;":PAGE:GLIS:DISP:LPAR ON"
OUTPUT @Hp4155;":PAGE:GLIS:DISP:LPAR?"
ENTER @Hp4155;A
Parameter Type Explanation
OFF or 0 boolean grid is off
ON or 1 boolean grid is on
Parameter Type Explanation
OFF or 0 boolean line parameters are not displayed
ON or 1 boolean line parameters are displayed