Agilent 4155C/4156C SCPI Command Reference, Edition 1 5-257
SCPI Commands
:PAGE:STRess[:CDEFinition]:SMU<n>:FUNCtion
:PAGE:STRess[:CDEFinition]:SMU<n>:FUNCtion
This command sets the function (FCTN) of SMU<n> for stress.
<n> is required to specify SMU number. Valid SMU numbers are SMU1 through
SMU6, depending on the configuration.
At *RST, this value is:
Syntax :PAGE:STRess [ :CDEFinition] :SMU<n>:FUNCtion SYNC | NSYNc
Parameter
Query response SYNC | NSYN | DIS <newline><^END>
If SMU is not used, the response data may be DIS (DISable).
Example OUTPUT @Hp4155;":PAGE:STR:SMU1:FUNC SYNC"
OUTPUT @Hp4155;":PAGE:STR:SMU1:FUNC?"
ENTER @Hp4155;A$
SMU FUNCTION
SMU1 NSYNC
SMU2 to 4 SYNC
SMU5 and SMU6 not defined
Parameter Type Explanation
SYNC character synchronous stress output
NSYNc character non-synchronous (non-stress output)