5-250 Agilent 4155C/4156C SCPI Command Reference, Edition 1
SCPI Commands
:PAGE:STRess[:CDEFinition]:ALL:DISable
:PAGE:STRess[:CDEFinition]:ALL:DISable
This command deletes the stress settings of all units (SMU,VSU,PGU,GNDU).
This command does not have query form.
Syntax :PAGE:STRess [ :CDEFinition] :ALL:DISable
Example OUTPUT @Hp4155;":PAGE:STR:ALL:DIS"
:PAGE:STRess[:CDEFinition]:COMMent
This command sets the USER DEFINED COMMENT for the stress group.
At *RST, this value is not defined.
Syntax :PAGE:STRess [ :CDEFinition] :COMMent comment
Parameter
Query response comment <newline><^END>
Example OUTPUT @Hp4155;":PAGE:STR:COMM 'PulseStress'"
OUTPUT @Hp4155;":PAGE:STR:COMM?"
ENTER @Hp4155;A$
:PAGE:STRess[:CDEFinition]:GNDU:DISable
This command disables the stress settings of GNDU.
This command does not have query form.
Syntax :PAGE:STRess [ :CDEFinition] :GNDU:DISable
Example OUTPUT @Hp4155;":PAGE:STR:GNDU:DIS"
Parameter Type Explanation
comment string String of up to 58 characters.