5-278 Agilent 4155C/4156C SCPI Command Reference, Edition 1
SCPI Commands
:PAGE:STRess:SETup:SSTop
OUTPUT @Hp4155;":PAGE:STR:SET:PULS:PGU1:WIDTH? MIN"
ENTER @Hp4155;A
:PAGE:STRess:SETup:SSTop
This command sets the stress stop mode for an abnormal status.
The abnormal statuses that can be detected are as follows:
• SMU reaches its compliance setting.
• Current of VSU exceeds ±100 mA.
• SMU or VSU oscillates.
• A/D converter overflow occurs.
• Average current of PGU exceeds ±100 mA.
At *RST, this value is OFF.
Syntax :PAGE:STRess:SETup:SSTop ABNormal | COMPliance | OFF
Parameter
Query response ABN | COMP | OFF <newline><^END>
Example OUTPUT @Hp4155;":PAGE:STR:SET:SST ABN"
OUTPUT @Hp4155;":PAGE:STR:SET:SST?"
ENTER @Hp4155;A$
Parameter Type Explanation
ABNormal character Stop stress if abnormal status is detected.
COMPliance character Stops stress only if some SMU reaches its
compliance setting.
OFF character Continue stress even if abnormal status is
detected.

Table of Contents

Other manuals for Agilent Technologies 4156C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4156C and is the answer not in the manual?

Agilent Technologies 4156C Specifications

General IconGeneral
BrandAgilent Technologies
Model4156C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals