:PAGE:MEASure:MSETup:ITIMe:LONG:APERture?
:PAGE:MEASure:MSETup:ITIMe:LONG:APERture?
This command returns the LONG integration time in terms of the APERTURE,
which means time.
The APERture value is related to NPLCycles by the following equation:
APERture = NPLCycles / selected line frequency
This command has query form only.
At *RST, this value is 16/selected line frequency.
Syntax :PAGE:MEASure:MSETup:ITIMe:LONG:APERture?
Query response aperture <newline><^END>
aperture is NR3 response data type.
Example OUTPUT @Hp4155;":PAGE:MEAS:MSET:ITIM:LONG:APER?"
ENTER @Hp4155;A
:PAGE:MEASure:MSETup:ITIMe:LONG[:NPLCycle]
This command sets the LONG integration time in terms of the Number of Power
Line Cycles. At *RST, this value is 16.
Syntax :PAGE:MEASure:MSETup:ITIMe:LONG [ :NPLCycle] nplcycle | MINimum |
MAXimum
Parameter
Query response nplcycle <newline><^END>
nplcycle is NR3 response data type.
Example OUTPUT @Hp4155;":PAGE:MEAS:MSET:ITIM:LONG 16"
OUTPUT @Hp4155;":PAGE:MEAS:MSET:ITIM:LONG MAX"
OUTPUT @Hp4155;":PAGE:MEAS:MSET:ITIM:LONG?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:MEAS:MSET:ITIM:LONG? MAX"
ENTER @Hp4155;A
Parameter Type Explanation
nplcycle numeric 2 to 100

Table of Contents

Other manuals for Agilent Technologies 4156C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4156C and is the answer not in the manual?

Agilent Technologies 4156C Specifications

General IconGeneral
BrandAgilent Technologies
Model4156C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals