Agilent 4155C/4156C SCPI Command Reference, Edition 1 5-25
SCPI Commands
:DISPlay:BRIGhtness
:DISPlay:BRIGhtness
This command controls the intensity of the display.
At *RST, this value is 0.843.
Syntax :DISPlay:BRIGhtness intensity | MINimum | MAXimum
Parameter
Query response intensity <newline><^END>
The data type of intensity is NR3 response data.
Example OUTPUT @Hp4155;":DISP:BRIG 0.5"
OUTPUT @Hp4155;":DISP:BRIG?"
ENTER @Hp4155;A
Parameter Type Explanation
intensity numeric 0.15 to 1. 1 is full intensity, and 0.15 is fully
blanked.