EasyManuals Logo

Agilent Technologies 4156C Command Reference

Agilent Technologies 4156C
424 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #296 background imageLoading...
Page #296 background image
5-214 Agilent 4155C/4156C SCPI Command Reference, Edition 1
SCPI Commands
:PAGE:MEASure:SAMPling:PERiod:AUTO
:PAGE:MEASure:SAMPling:PERiod:AUTO
This command sets the TOTAL SAMPLING TIME for sampling measurement.
This command is available only for LINear mode.
If you set :PAGE:MEAS:SAMP:PER total_samp_time, then AUTO OFF is set.
Syntax :PAGE:MEASure:SAMPling:PERiod:AUTO OFF | ON | 0 | 1
Parameter
Query response 0 | 1 <newline><^END>
Example OUTPUT @Hp4155;":PAGE:MEAS:SAMP:PER:AUTO ON"
OUTPUT @Hp4155;":PAGE:MEAS:SAMP:PER:AUTO?"
ENTER @Hp4155;A
:PAGE:MEASure:SAMPling:POINts
This command sets the NUMBER OF SAMPLES for sampling measurement.
Syntax :PAGE:MEASure:SAMPling:POINts no_of_samples | MINimum | MAXimum
Parameter
Query response no_of_samples <newline><^END>
no_of_samples is NR1 response data type.
Parameter Type Explanation
OFF or 0 boolean :PAGE:MEAS:SAMP:PER setting is effective.
ON or 1 boolean auto (disables total sampling time stop event, and
enables the number of samples stop event.)
Parameter Type Explanation
no_of_samples numeric 1 to 10001

Table of Contents

Other manuals for Agilent Technologies 4156C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4156C and is the answer not in the manual?

Agilent Technologies 4156C Specifications