Agilent 4155C/4156C SCPI Command Reference, Edition 1 5-129
SCPI Commands
:PAGE:DISPlay[:SETup]:LIST[:SELect]
:PAGE:DISPlay[:SETup]:LIST[:SELect]
This command selects the variable names for LIST display.
You can display a maximum of eight variables in the list. If this limit is exceeded,
Cannot define more than 8 data vars in lists. error (-183) is
generated.
On the DISPLAY:DISPLAY SETUP page, this command fills the blank LIST fields
in the specified order.
At *RST, no variable is selected.
Syntax :PAGE:DISPlay [ :SETup] :LIST [ :SELect] var_name { ,var_name}
Parameter
Query response var_name <newline><^END>
var_name is a list of variable names separated by commas, and is string response
data, but does not contain double quote characters at the beginning and end of the
string.
Example OUTPUT @Hp4155;":PAGE:DISP:LIST 'VD','VG'"
OUTPUT @Hp4155;":PAGE:DISP:LIST?"
ENTER @Hp4155;A$
:PAGE:DISPlay[:SETup][:MENU]
This command changes the present display page to DISPLAY: DISPLAY SETUP.
This command does not have query form.
Syntax :PAGE:DISPlay [ :SETup] [ :MENU]
Example OUTPUT @Hp4155;":PAGE:DISP"
Parameter Type Explanation
var_name string data variable name

Table of Contents

Other manuals for Agilent Technologies 4156C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4156C and is the answer not in the manual?

Agilent Technologies 4156C Specifications

General IconGeneral
BrandAgilent Technologies
Model4156C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals