Agilent 4155C/4156C SCPI Command Reference, Edition 1 5-233
SCPI Commands
:PAGE:MEASure[:SWEep]:VAR2:PCOMpliance:STATe
Parameter
Query response pcompliance <newline><^END>
pcompliance is NR3 response data type.
Example OUTPUT @Hp4155;":PAGE:MEAS:VAR2:PCOM 0.5"
OUTPUT @Hp4155;":PAGE:MEAS:VAR2:PCOM MAX"
OUTPUT @Hp4155;":PAGE:MEAS:VAR2:PCOM?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:MEAS:VAR2:PCOM? MAX"
ENTER @Hp4155;A
:PAGE:MEASure[:SWEep]:VAR2:PCOMpliance:STA
Te
This command sets the power compliance of VAR2 to disable.
You use this command only if there is an SMU whose function (FCTN) is VAR2.
If the unit type of VAR2 is VSU, this parameter is ignored.
If power compliance value is set by PCOMpliance pcompliance, the query returns
1.
At *RST, this value is OFF.
Syntax :PAGE:MEASure [ :SWEep] :VAR2:PCOMpliance:STATe OFF | 0
Parameter
Query response 0 | 1 <newline><^END>
Example OUTPUT @Hp4155;":PAGE:MEAS:VAR2:PCOM:STATE OFF"
OUTPUT @Hp4155;":PAGE:MEAS:VAR2:PCOM:STATE?"
ENTER @Hp4155;A
Parameter Type Explanation
pcompliance numeric 1E-3 to 20. The range of this value depends on
SMU type of VAR2.
Parameter Type Explanation
OFF or 0 boolean power compliance is disabled