5-276 Agilent 4155C/4156C SCPI Command Reference, Edition 1
SCPI Commands
:PAGE:STRess:SETup:PULSe:PGU<n>:PERiod
:PAGE:STRess:SETup:PULSe:PGU<n>:PERiod
This command sets the pulse PERIOD of PGU<n> for stress.
<n> is required to specify PGU number. Valid PGU numbers are PGU1 through
PGU2.
PGU1:PERiod and PGU2:PERiod are always set to the same value.
The mode of the specified PGU must be VPULse.
At *RST, this value is 10 ms.
Syntax :PAGE:STRess:SETup:PULSe:PGU<n>:PERiod period | MINimum | MAXimum
Parameter
Query response period <newline><^END>
period is NR3 response data type.
Example OUTPUT @Hp4155;":PAGE:STR:SET:PULS:PGU1:PER 0.01"
OUTPUT @Hp4155;":PAGE:STR:SET:PULS:PGU1:PER MAX"
OUTPUT @Hp4155;":PAGE:STR:SET:PULS:PGU1:PER?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:STR:SET:PULS:PGU1:PER? MAX"
ENTER @Hp4155;A
:PAGE:STRess:SETup:PULSe:PGU<n>:TRAiling
This command sets the trailing-edge transition time (TRAILING TIME) of
PGU<n> for stress.
<n> is required to specify PGU number. Valid PGU numbers are PGU1 through
PGU2.
The mode of the specified PGU must be VPULse.
At *RST, this value is 100 ns.
Syntax :PAGE:STRess:SETup:PULSe:PGU<n>:TRAiling trailing_time | MINimum |
MAXimum
Parameter Type Explanation
period numeric 2E-6 to 10 (s)

Table of Contents

Other manuals for Agilent Technologies 4156C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4156C and is the answer not in the manual?

Agilent Technologies 4156C Specifications

General IconGeneral
BrandAgilent Technologies
Model4156C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals