5-174 Agilent 4155C/4156C SCPI Command Reference, Edition 1
SCPI Commands
:PAGE:MEASure:OSEQuence[:MENU]
:PAGE:MEASure:OSEQuence[:MENU]
This command changes the present display page to MEASURE: OUTPUT
SEQUENCE page. This command does not have query form.
Syntax :PAGE:MEASure:OSEQuence [ :MENU]
Example OUTPUT @Hp4155;":PAGE:MEAS:OSEQ"
:PAGE:MEASure:OSEQuence:MODE
This command selects the output sequence mode for sampling measurement.
You use this command only if the measurement mode is sampling.
At *RST, this value is SIMultaneous for sampling measurement.
Syntax :PAGE:MEASure:OSEQuence:MODE SEQuential | SIMultaneous
Parameter
Query response SEQ | SIM <newline><^END>
Example OUTPUT @Hp4155;":PAGE:MEAS:OSEQ:MODE SIM"
OUTPUT @Hp4155;":PAGE:MEAS:OSEQ:MODE?"
ENTER @Hp4155;A$
:PAGE:MEASure:OSEQuence:OSEQuence
This command sets the OUTPUT SEQUENCE.
The source channels output in the order specified by the OUTPUT SEQUENCE.
Unspecified units output in default order after the specified units.
At *RST, the output sequence is SMU1-SMU2-SMU3-SMU4-SMU5-SMU6-
VSU1-VSU2-PGU1-PGU2.
Syntax :PAGE:MEASure:OSEQuence:OSEQuence unit { , unit }| DEFault
Parameter Type Explanation
SEQuential character sequential output
SIMultaneous character simultaneous output