EasyManua.ls Logo

Agilent Technologies 4156C

Agilent Technologies 4156C
424 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Agilent 4155C/4156C SCPI Command Reference, Edition 1 5-223
SCPI Commands
:PAGE:MEASure[:SWEep][:MENU]
OUTPUT @Hp4155;":PAGE:MEAS:HTIM?"
ENTER @Hp4155;A
:PAGE:MEASure[:SWEep][:MENU]
This command changes the present display page to MEASure: SWEEP SETUP.
This command does not have query form.
Syntax :PAGE:MEASure [ :SWEep] [ :MENU]
Example OUTPUT @Hp4155;":PAGE:MEAS:SWE"
:PAGE:MEASure[:SWEep]:PULSe:BASE
This command sets the pulse BASE of SMU.
You use this command only if an SMU is set to VPULse or IPULse mode.
Syntax :PAGE:MEASure [ :SWEep] :PULSe:BASE base | MINimum | MAXimum
Parameter
Query response base <newline><^END>
base is NR3 response data type.
Example OUTPUT @Hp4155;":PAGE:MEAS:PULS:BASE 1.5"
OUTPUT @Hp4155;":PAGE:MEAS:PULS:BASE MIN"
OUTPUT @Hp4155;":PAGE:MEAS:PULS:BASE?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:MEAS:PULS:BASE? MIN"
ENTER @Hp4155;A
Parameter Type Explanation
base numeric -200 to 200 V or -1 to 1 A. The range of this value
depends on the type of SMU.

Table of Contents

Other manuals for Agilent Technologies 4156C

Related product manuals