Agilent 4155C/4156C SCPI Command Reference, Edition 1 5-223
SCPI Commands
:PAGE:MEASure[:SWEep][:MENU]
OUTPUT @Hp4155;":PAGE:MEAS:HTIM?"
ENTER @Hp4155;A
:PAGE:MEASure[:SWEep][:MENU]
This command changes the present display page to MEASure: SWEEP SETUP.
This command does not have query form.
Syntax :PAGE:MEASure [ :SWEep] [ :MENU]
Example OUTPUT @Hp4155;":PAGE:MEAS:SWE"
:PAGE:MEASure[:SWEep]:PULSe:BASE
This command sets the pulse BASE of SMU.
You use this command only if an SMU is set to VPULse or IPULse mode.
Syntax :PAGE:MEASure [ :SWEep] :PULSe:BASE base | MINimum | MAXimum
Parameter
Query response base <newline><^END>
base is NR3 response data type.
Example OUTPUT @Hp4155;":PAGE:MEAS:PULS:BASE 1.5"
OUTPUT @Hp4155;":PAGE:MEAS:PULS:BASE MIN"
OUTPUT @Hp4155;":PAGE:MEAS:PULS:BASE?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:MEAS:PULS:BASE? MIN"
ENTER @Hp4155;A
Parameter Type Explanation
base numeric -200 to 200 V or -1 to 1 A. The range of this value
depends on the type of SMU.

Table of Contents

Other manuals for Agilent Technologies 4156C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4156C and is the answer not in the manual?

Agilent Technologies 4156C Specifications

General IconGeneral
BrandAgilent Technologies
Model4156C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals