Agilent 4155C/4156C SCPI Command Reference, Edition 1 5-171
SCPI Commands
:PAGE:MEASure:MSETup:SMU<n>|:VMU<n>:RANGe:MODE
:PAGE:MEASure:MSETup:SMU<n>|:VMU<n>:RAN
Ge:MODE
This command selects the ranging MODE of SMU<n> or VMU<n>.
<n> is required to specify SMU or VMU number. Valid numbers are SMU1 through
SMU6 (depending on the configuration) and VMU1 through VMU2.
The specified SMU or VMU must not be DISable.
Syntax :PAGE:MEASure:MSETup:SMU<n> | :VMU<n>:RANGe:MODE AUTO | FIXed |
LIMited
Parameter
Query response AUTO | FIX | LIM <newline><^END>
Example OUTPUT @Hp4155;":PAGE:MEAS:MSET:SMU1:RANG:MODE AUTO"
OUTPUT @Hp4155;":PAGE:MEAS:MSET:SMU1:RANG:MODE?"
ENTER @Hp4155;A$
Parameter Type Explanation
AUTO character auto range mode
FIXed character fixed range mode
LIMited character limited auto range mode